MHK ISDB/Sensors/Reson TC4035 Hydrophone

Jump to: navigation, search

Reson TC4035 Hydrophone

Manufacturer: Teledyne RESON
Measurement Category: Acoustic Noise
Operating Environment: Underwater

Communication Protocol: Voltage
Operating Mode: Free Streaming
Power Supply Type: VDC
Power Specification: 10-24 VDC; 15 mA at 12 VDC


The TC 4035 is a miniature probe hydrophone specifically designed as a standard reference for sound measurements in the frequnecy range 100-500 kHz. Incoperates a 10 dB low noise preamplifier. Each hydrophone is individulaly calibrated.

Sensor Measurements

  • Sound Pressure Level (Sound Pressure)

Measurement Specifications
Frequency response: 1 00 kHz to 500 kHz Nominal Sensitivity: -214dB (+-2 dB) re 1V/uPa (w/o preamp)

Guides and Manuals

Product Web Page:

Notes, Considerations & Recommendations

No notes have been provided. If you have worked with this sensor, please contribute your notes.

User Experience

Contribute your experience